ECG | 12-CHANNEL ECG MACHINE | MEDLI-12C CHANNEL(WITH TOUCH SCREEN)

ECG Machine- Medico Lifeline

MEDLI-12C CHANNEL (WITH TOUCH SCREEN)

Functional Characteristics

  • 12 leads ECG synchronous sampling system.
  • Thermal printer with 216mm print paper, synchronization paper.
  • 10.2 -inch color touch screen, both push-button, and touch operation.
  • Automatic measurement, calculation analysis, waveform freezing.
  • Support USB flash disk and SD memory.
  • Data transmitting to PC via RS232/USB.
  • Patient information can be input.
  • Historical data can be reviewed and printed.
  • Built-in rechargeable lion battery, AC/DC power conversion.
  • Sleep mode to save energy and extended LCD life.
  • Portable design with hidden handle convenient to carry.
  • Leadoff and lack of paper detection function. 7 level printing thick can be chosen.
  • Adapt to 110-230v, 50/60 Hz AC power supply.

Product Features 

Touch mode, One key operation, Quick input, Unlimited storage, Simple review

Input wayFloating and Defibrillation protection
Patient leak current< 10uA
Input Impedance
Frequency response
Time constant
5ohm
0.05Hz-150Hz(3dB)
Time constant >3.2s
CMRR>60dB, >100dB(Adding Filter)
EMG interference filter
Recording way
Specification of recording paper
35Hz(-3 dB)
Thermal printing system
210mm(W)*20m(L) high-speed thermal paper
Paper SpeedAuto record: 25mm/s, 50mm/s, error: ±5%
Rhythm record: 25mm/s, 50mm/s, error: ±5%
Manual record: 5mtn/s, 6.25mm/s, 10mm/s, 12.5mtn/s, 25mm/s, 50mm/s, error: ±5%
Sensitivity Selections

Auto record
2.5,4,10,20,40 mm/mV, error: ±5%, Standard sensitivity is 10mm/mV ± 0.2 mm/mV
Record setup according to auto record format and mode, automatically changing leads, measuring and analysing.
Rhythm record

Manual record
Record setup according to rhythm record format and mode, automatically measuring and analysing.
Record setup according to manual record format, manually changing leads.
Measurement parameters
Product safety type
Noise level
HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, O-Tc, P A*is, ORS A*is, T A*is, R(V5) S(V1) R(V5)+S(V1)
Class I CF applied part; there is defibrillation and pacing protection circuit in it. = 15uVn.p
Size340mm(L)*320mm(W)*85mm(H)
Net Weight3.2 Kg

MEDLI-6C (WITH TOUCH SCREEN)